Fluke 9009 Dual Dry Block Temperature Calibrator
Product Description
The Model 9009 includes two independently controlled temperature blocks. The hot block provides temperatures from 50 °C to 350 °C, while the cold block covers the range –15 °C to 110 °C. Each block is controlled by a precision Hart Scientific temperature controller. These aren’t some off-the-shelf controllers we glued into a box. These are Hart Scientific controllers from the leading temperature company in the world.
Each temperature block includes two wells with removable inserts. You can calibrate four probes at once, or you can calibrate two probes at the same time with an external reference (like Hart’s Model 1521 LLK Thermometer), or you can use the two temperature wells to get quick “zero" and “span" references for transmitter calibrations.
Overview
-
Double your productivity or cut your calibration time in half
- Temperatures from –15 °C to 350 °C in one unit
- Two wells in each block for simultaneous comparison calibrations
- Rugged, lightweight, water-resistant enclosure
You’ve been asking for it and now we’re making it for you. Hart’s new Model 9009 Industrial Dual-Block Calibrator lets you calibrate at hot and cold temperatures at the same time. Double your productivity or cut your calibration time in half—either way you look at it, your in-field temperature calibrations just got easier.
Specifications
General Specifications | ||
Hot block | Cold block | |
Range | 50°C to 350°C (122°F to 662°F) | –15°C to 110°C (5°F to 230°F) (–8°C [18°F] with hot block at 350°C [662°F]) |
Accuracy | ±0.6°C | ±0.2°C |
Stability | ±0.05°C | |
Well-to-well uniformity | ±0.1°C | |
Display resolution | 0.1° | |
Heating times | 30 minutes from 25°C to 350°C | 15 minutes from 25°C to 110°C |
Cooling times | 40 minutes from 350°C to 100°C | 16 minutes from 25°C to –15°C |
Stabilization times | 8 minutes | |
Well depth | 4 in (102 mm) | |
Removable inserts | Two 6.4 mm (1/4 in) and two 4.8 mm (3/16 in) inserts included; see ordering information for other available inserts | |
Computer interface | RS-232 included | |
Power | 115 V AC (±10 %), 3 A, or 230 V AC (±10 %), 2 A, specify, 50/60 Hz, 280 W |
|
Size (H x W x D) | 178 x 267 x 248 mm (7 x 10.5 x 9.75 in) | |
Weight | 4.5 kg (10 lb) | |
NIST-traceable calibration | Data at 50°C, 100°C, 150°C, 200°C, 250°C, 300°C, and 350°C | Data at –8°C, 0°C, 25°C, 50°C, 75°C, 100°C, and 110°C |