Fluke 789 ProcessMeter™
Product Description
Fluke 789 ProcessMeter doubles your troubleshooting power and is Fluke Connect-enabled
The Fluke 789 ProcessMeter is the ultimate troubleshooting tool for electricians and instrumentation professionals. It combines the functionality of a loop calibrator with the power of a digital multimeter, and the wireless capabilities of Fluke Connect connectivity with optional IR3000FC module. The 789 has a display that's not only twice as large as the Fluke 787B, but also twice as bright with two levels of backlighting. And its built-in, selectable 250 ohm HART® resistor eliminates the need to carry a separate resistor.
The 789 also has a 24-Volt loop supply, which reduces the need to carry a separate power supply when doing offline transmitter testing. You can add an optional temperature measurement module and the wireless data logging capabilities of the Fluke Connect™ mobile app to log data wirelessly to your smart phone and review it from a remote location. So now process technicians can do a lot more while carrying a lot less.
Overview
Key features
- Ultimate troubleshooting tool for electricians and instrumentation professionals
- Combines the functionality of a loop calibrator with the power of a precision 1000 V, 440 mA True-RMS digital multimeter
- Optional IR3000FC module for wireless data logging capabilities via Fluke Connect® mobile App
- Includes 24 V loop power supply
- Features a very large dual display with enhanced two-level backlight brightness settings
Specifications
Specifications | ||
Current measurement | Range | 0-1 A AC or DC |
Resolution | 1 mA | |
Accuracy | 0.2% + 2 LSD (DC) | |
Range | 0-30 mA | |
Resolution | 0.001 mA | |
Accuracy | 0.05% + 2 lSD | |
Current sourcing | Range | 0-20 mA or 4-20 mA |
Resolution | 0.05% of span | |
Drive capability | 24 V Compliance or 1200 Ω @ 20 mA | |
Frequency | to 19.999 kHz, 0.005% + 1 LSD | |
Continuity | Beeps for resistance < 100 Ω | |
Voltage measurement | Range | 0-1000 V AC (true-rms) or DC |
Resolution | 0.1 mV to 1.0 V | |
Accuracy | 0.1% Rdg + 1 LSD (VDC) | |
Resistance measurement | to 40 MΩ, 0.2% + 1 LSD | |
Diode test | 2.0 V shoes diode voltage drop | |
Display current and % of span | Yes | |
Auto Step, Auto Ramp | Yes | |
Environmental Specifications | ||
Operating temperature | -20°C to 55°C | |
Storage temperature | -40°C to 60°C | |
Humidity (without condensation) | 95% up to 30°C 75% up to 40°C 45% up to 50°C 35% up to 55°C |
|
Temperature coefficient | 0.05 x (specified accuracy) per °C (for temperatures < 18°C or > 28°C) | |
Safety Specifications | ||
Electrical safety | Designed in accordance with IEC 1010-1, ANSI/ISA S82.01-1994 and CAN/CSA C22.2 No 1010.1-92 | |
Overvoltage category | IEC 1010-1 CAT III 1000 V IEC 1010-1 CAT IV 600 V |
|
Maximum voltage | 1000 V RMS | |
Mechanical and General Specifications | ||
Size (H x W x L) | 50 x 100 x 203 mm (1.25 x 3.41 x 7.35 in) | |
Weight | 610 g (1.6 lbs) | |
Batteries | 4 AA alkaline batteries | |
Warranty | 3 years | |
Battery life | 140 hours typical (measurement), 10 hours typical (sourcing 20 mA) | |
Shock and vibration | Random, 2 g, 5-500 Hz 1 meter drop test |
What's in the Box
Includes:
- AC175 Alligator Clips
- Four (4) AA 1.5 V alkaline batteries (installed)
- TL71 safety-designed test lead set